Sympatec PAQXOS Export File Format Reference

This pipeline ingests the per-measurement exports produced by Sympatec's PAQXOS laser-diffraction software. This document records the raw layout so ingest (Step 0, extract_run) can be implemented and maintained without depending on undocumented institutional knowledge. The layout is generated through PAQXOS's normal Report Template Editor and parsed from the resulting RTF export. The pipeline does not access PAQXOS's internal databases or executables.

The report-template body used to request optical concentration plus reference and measured ring intensities is included at templates/paqxos_raw_intensity_report_template.txt. It was reconstructed from the lab's archived print-to-PDF copy of the working template (SHA-256 177fdf18f15c7bc42779b4509ec953a309ecad8b77544b2e019fe55ed562a34c). Paste the text into an empty PAQXOS report template and validate it with the editor's report preview before operational use.

PAQXOS 5.1 Operating Instructions document the Report Template Editor and @I.REF(N). The other substitutions in the bundled template are evidenced by the archived working template and rendered exports. In particular, @I.NORM(N) was not found by text search of the PAQXOS 5.1 or WINDOX 5.6 command-reference PDFs reviewed for this release.

Container

  • Exports are RTF (.rtf); one file per run (measurement or blank).
  • De-RTF'd to plain text via striprtf.rtf_to_text.

Per-frame structure

A file is a sequence of frames. Each frame:

Measurement Name:  Compound X dissolution measurement
Measurement Time:   2026-01-01 15:49:38
Report Generated on: 2026-06-09 17:51:36
Optical Concentration: 1.70
Channel, Ref Value, Measured Value
1, 1.80, 1.40
2, 0.68, 0.64
... (31 channels)
Marker / row Meaning
Measurement Name: run label (constant within a file)
Measurement Time: YYYY-MM-DD HH:MM:SS frame timestamp
Optical Concentration: <float> optical concentration (Copt), as reported by PAQXOS
Channel, Ref Value, Measured Value column header
<ch>, <ref>, <measured> per detector ring: index, reference, measured

Parser contracts

  1. Two intensity columns. Measured is the per-frame signal (analysis target); Ref is the instrument's stored reference spectrum. Capture bothRef feeds optional static-baseline subtraction (I_bgsub = I − ref) and the reference-variation check. The adapter retains the earliest valid frame's stored-reference vector and flags variation across retained frames; variation does not automatically exclude a frame.
  2. Frame structure is validated against detector identifiers. A PAQXOS instrument profile may declare numeric channel_ids in strictly increasing order. A frame is retained only when its timestamp is parseable and every declared identifier occurs exactly once. Missing, additional, substituted, or duplicate identifiers are counted by exclusion reason. Without declared identifiers, the adapter infers a uniquely most frequent exact set and fails closed when equally supported sets are ambiguous.
  3. Frames may be listed newest-first. The parser sorts by timestamp ascending and records whether reverse document order was detected.

Measured is not background-subtracted: in a blank, Measured ≈ Ref ≈ 0.3–0.6 (not zero). Measured − Ref is the drug-attributable scattering, offered as a derived I_bgsub; Measured is the primary signal.

The current parser treats each new Measurement Time: marker as the start of a frame and sorts parsed timestamps chronologically. It does not depend on @PAR(1) as a structural separator. In the PAQXOS 5.1 manual, @PAR(1) means the first user-defined measurement parameter; it is retained in the archived template body for fidelity but is not required by the parser or by intensity extraction. Missing Copt, reference variation, reverse document order, and interframe gaps are reported as run-level flags, not additional automatic frame exclusions.

Output contract

The PAQXOS adapter returns the instrument-neutral Run contract described in data_contracts.md. Existing analyses can still access the legacy names I, ref, copt, t_min, t0, and channels through compatibility properties. These names are not requirements for other adapters.

CSV mirror

dfm-ingest / extract_run(emit_csv=True) writes one tidy CSV per run — frame, time_iso, t_min, copt, I_ch1 … I_ch31 — plus a <run>_meta.json sidecar carrying the earliest retained frame's stored-reference vector and run-level flags.

Open questions

  • Validation of the nominal manual-derived channel annuli and any size-specific interpretation using independent monodisperse standards or controlled mixtures. A channel is not a particle-size bin.
  • Whether multiple lenses/configs appear across the dataset.